The Flow Cytometry Shared Resource Core features a comprehensive suite of instrumentation to facilitate your research needs. These include:

Canto - BD FACSCanto™ II Analyzer

The BD FACSCanto™ II located in 4-001 MMRB is equipped with BD FACS carousel loader and 2 lasers (blue: 488 nm, red: 633 nm) allowing the use of up to 7 color simultaneous detection. The BD FACSCANTO II™ analyzer is equipped with a uses FACSDIVA software for acquisition and analysis. DeNovo FCSExpress software is available for remote analysis.
FACSCanto

Laser, detector/filter configuration ofthe BD FACSCanto™ II analyzer:

Laser

LP Mirror

BP Filter

Intended Dye

488 nm

735 nm

780/60 nm

PE-Cy7

 

655 nm

670 nm

PerCP, PerCP-Cy5.5, PE-Cy5

 

595 nm

610/20 nm

PE-TexasRed

 

556 nm

585/42 nm

PE

 

502 nm

530/30 nm

FITC, Alexa 488, GFP, CFSE

 

Blank

488/10and pinhole

SSC

633 nm

735 nm

780/60 nm

APC-Cy7

 

Blank

660/20 nm

APC, Alexa 647

Optional

685 nm

710/40 nm

Alexa700

 

Aria-BD FACSAria™ II High-Speed Cell Sorter

This cell sorter located in 4-001 MMRB is equipped with three lasers (405nm, 488nm, 633nm) allowing the use of 9 color simultaneous detection. Capabilities include various sort nozzle sizes, four-way sorting and single cell cloning.  It is equipped with the temperature control option for the sort sample and collection vials and the BD Aerosol Management System for sorting of potentially biohazard us samples. The sorter uses FACSDIVA software for acquisition/analysis. DeNovo FCSExpress software is available for remote and offline analysis.Please contact core personnel with questions or to schedule a sort.

BSC Aria- BD FACSAria™ II High-Speed Cell Sorter

FACSAria

Our second cell sorter located in 4-001 MMRB is placed inside a Baker BioProtect Hood for sorting of potentially biohazardous samples. 

This sorter is equipped with three lasers (488nm, 633nm, and 375nm) allowing 9 color simultaneous detection. Capabilities include various sort nozzle sizes, four-way sorting and single cell cloning.  It is equipped with the temperature control option for the sort sample and collection vials, and the BD Aerosol Management System for additional containment. The sorter uses FACSDIVA software for acquisition/analysis. DeNovo FCSExpress software is available for remote and offline analysis. Please contact core personnel with questions or to schedule a sort

FC500

FC-500

The FC500 located in 327 KMSB is equipped with 2 lasers (488nm, 633nm) allowing 5 color simultaneous detection. This analyzer uses CXP software for acquisition/analysis. DeNovo FCSExpress software is available for remote and offline analysis.

 

XL - Beckman Coulter Epics XL-MCL

In addition to our front-line instruments, the flow core maintains an Elite XL-MCL single-laser flow cytometer suitable for the majority of common flow cytometry applications, including immunofluorescence and DNA analysis, cell viability studies, and mitochondrial analyses. The XL located in 327 KMSB is equipped with 1 laser (488nm) allowing 4 color simultaneous detection. This analyzer uses EXPO32 software for acquisition/analysis. DeNovo FCSExpress software is available for remote and offline analysis.

Elite XL-MCL

TopCount

Microplate luminesence counting services are available with our PerkinElmer TopCount microplate reader.

 

TopCount

BiaCore T200 Surface Plasmon Resonance